510(k) Data Aggregation

    K Number
    K193447

    Validate with FDA (Live)

    Device Name
    Vista Dyes
    Date Cleared
    2020-03-11

    (89 days)

    Product Code
    Regulation Number
    872.1740
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A visual aid for the identification of carious dentin.

    Device Description

    Vista Dyes are colored agents that when applied to the suspected carious areas of the tooth stains the carious dentin. The method of operation and placement of Vista Dyes are similar to that of the primary predicate and reference device, Caries Finder (K955445) and Pulpdent Snoop (K964430), respectively. A drop of the dye is applied to the carious dentine, rinsed with water and air dispersed. The stained carious tissue is then removed with a low speed rotary bur. The process may be repeated until there are no more stainable tissues in the carious cavity.

    AI/ML Overview

    The device described is "Vista Dyes", a caries detection device. Based on the provided text, there is no detailed study provided that presents specific acceptance criteria and device performance metrics in a quantitative manner. The submission primarily relies on demonstrating substantial equivalence to predicate devices (Caries Finder and Pulpdent Snoop) based on identical indications for use, technological characteristics, and methods of application, rather than presenting a performance study with specific acceptance criteria.

    The document states: "Clinical performance is not deemed necessary to support the substantial equivalence of the proposed device." Therefore, there is no information about acceptance criteria, reported device performance, sample size, ground truth establishment, or expert involvement for a performance study.

    However, based on the information provided, we can infer some aspects of what would typically be considered "acceptance criteria" through the comparative analysis with predicate devices.

    Here's an analysis of the provided information in response to your questions:

    1. A table of acceptance criteria and the reported device performance

    No explicit quantitative acceptance criteria or reported device performance (e.g., sensitivity, specificity) were provided in the document. The substantial equivalence argument serves as the "acceptance criteria" here, meaning that the device's characteristics must be sufficiently similar to the predicate devices to not raise new questions of safety or effectiveness.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable. No clinical or performance test set was described.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. No clinical or performance test set requiring expert ground truth was described.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. No clinical or performance test set was described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a dye, not an AI-assisted device, and no MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a dye, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. No performance study requiring ground truth was conducted or described. The safety and effectiveness are established by substantial equivalence to legally marketed predicate devices, implying that the known performance characteristics of the predicate devices implicitly serve as a "ground truth" for what is acceptable.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device, so there is no training set.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/ML device, so there is no training set or associated ground truth establishment.

    Summary of Acceptance Criteria and Performance (Inferred from Substantial Equivalence):

    Acceptance Criteria CategoryReported Device Performance / Justification
    Indications for UseIdentical to predicate device (Caries Finder): "A visual aid for the identification of infected carious dentin."
    Mechanism of ActionIdentical to predicate: "Visible staining of carious dentin."
    Material Composition- Red & Green Dyes: 99% propylene glycol, 1% dye (acid red 52 or FD&C green). Identical to predicate.- Blue Dye: 99% distilled water, 1% methylene blue dye. Differs from primary predicate (which uses propylene glycol) but is equivalent to reference device (Pulpdent Snoop).
    Physical PropertiesColored aqueous liquid. Identical to predicate.
    Chemical Properties1% dye concentration. Identical to predicate.
    Intended UseSame target population and anatomical site. Identical to predicate.
    Packaging Configuration- Dropper bottles: Identical to predicate.- Prefilled 1.2mL syringes: Additional configuration, deemed not to raise new concerns as it simplifies application and is common in dentistry.
    SterilityNon-sterile. Identical to predicate.
    Shelf-Life3 years. Based on accelerated testing and ongoing real-time aging. Identical to predicate.
    Prescription/OTC UsePrescription use only. Identical to predicate.
    BiocompatibilityConfirmed via literature review, LD50 analysis of constituents, existing FDA data, and longstanding use of predicate without significant adverse events (MAUDE database).
    Safety and EffectivenessDemonstrated through substantial equivalence to legally marketed predicate devices, with no new questions of safety or effectiveness raised by minor differences.

    In essence, the "study" for Vista Dyes was a substantial equivalence comparison demonstrating that its characteristics are similar enough to existing, legally marketed devices that it does not require new clinical performance data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193389

    Validate with FDA (Live)

    Device Name
    Vista Clear
    Date Cleared
    2020-03-04

    (89 days)

    Product Code
    Regulation Number
    N/A
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Vista Clear is intended for sulcus retraction prior to impression making and to control bleeding and gingival oozing in restorative and operative dentistry used with gingival retraction cord. Vista Clear facilitates the insertion of the cord into the sulcus.

    Device Description

    Vista Clear is used to facilitate sulcus retraction prior to taking a dental impression of a tooth. This entails placement of the device into the sulcus which provides physical displacement of the gingival tissue from the tooth. If using a gingival retraction cord, the subject device facilitates the insertion of the cord into the sulcus while also creating a physical barrier to prevent gingival bleeding and oozing from affecting restorative and tissue management procedures.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called "Vista Clear." This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than undergoing extensive clinical trials or performance testing against specific acceptance criteria for efficacy in the same way a novel device might.

    Therefore, the document does not contain information on "acceptance criteria" and "study that proves the device meets the acceptance criteria" in the context of clinical performance or diagnostic accuracy. Instead, the "device performance" described relates to non-clinical testing demonstrating equivalence to a predicate device.

    Here's an analysis based on the provided text, addressing your questions to the extent possible, and highlighting what's not present:

    Key Takeaway: This 510(k) summary is for a device with a clear physical/chemical function (sulcus retraction, bleeding control) and claims substantial equivalence based on technological characteristics and non-clinical testing, not clinical performance or diagnostic accuracy with human interpretation.


    1. A table of acceptance criteria and the reported device performance

    The concept of "acceptance criteria" in this 510(k) largely revolves around demonstrating substantial equivalence to the predicate device and meeting relevant non-clinical performance standards. There are no explicit quantitative clinical performance acceptance criteria (e.g., sensitivity, specificity, AUC) stated in the document, as it's not a diagnostic device.

    Reported Device Performance (from "Non-Clinical Performance Testing and Compliance"):

    Acceptance Criteria (Implied by Equivalence & Standards)Reported Device Performance (Vista Clear)
    Technological Characteristics Substantial Equivalence:
    - Same Intended Use: Sulcus retraction and bleeding/gingival oozing control.- Meets. Nearly identical indications for use as the predicate device.
    - Same Common Name / Product Code Classification (MVL).- Meets.
    - Same Recommended Contact Time (1-3 minutes).- Meets.
    - Same Form/Mechanism: Aqueous gel for physical retraction.- Meets. "identical to the predicate device as both products are aqueous gels which aid in the physical retraction of gingival tissue."
    - Same Target Population & Anatomical Site: Healthcare professionals, oral cavity.- Meets.
    - Same Prescription Use classification.- Meets.
    - Same Packaging/Configuration.- Meets. "offered in the same configurations as the predicate device (i.e. prefilled syringes with applicator tips, or bulk syringes with unfilled smaller syringes and applicator tips)."
    - Similar Chemical Characteristics (Aluminum chloride based).- Meets. Contains 26.6% aluminum chloride hexahydrate vs. predicate's 25% aluminum chloride. Stated this difference "does not raise any safety or efficacy concerns as they are analogous materials."
    - Similar pH (Implied by "identical technological characteristics").- Meets. "Both medical devices have an identical pH."
    - Non-Sterile classification.- Meets.
    Non-Clinical Performance Acceptance Criteria:
    - Analytical Testing: Verified manufacturing and comparable results to predicate.- Meets. "Results from testing are commensurate with the predicate device, supporting substantial equivalence..."
    - Cytotoxicity Testing (ISO 10993-1): No significant cytotoxicity, comparable to predicate.- Meets. "Vista Clear exhibited the same cytotoxicity result as the predicate device (ViscoStat Clear) sold for the same intended use." This "confirms that the subject device is substantially equivalent to the predicate device."
    - Microbiological Testing: Mitigation of contamination risks, support shelf stability.- Meets. "Vista Clear exhibits bactericidal properties." This "helps to support shelf stability and multiple use of non-patient contacting materials." Note: Not claiming a "bactericidal effect" on the patient, but rather internal contamination control.
    - Shelf-Life Testing: Demonstrate stability for stated shelf-life (24 months).- Meets. "Based on accelerated testing, a shelf life of two years is supported for Vista Clear." (Predicate has 42 months, but this difference is deemed acceptable with appropriate labeling.) Real-time aging is ongoing.
    - Transit Testing: Packaging withstands simulated transit conditions without negative effect on product.- Meets. "Moreover, the products performed satisfactory post-transit. which confirms that transit did not have a negative effect on the products themselves."
    - Compliance with applicable standards (e.g., ISO 10993-1, ISO 14971, ISO 594-1/2 for Luer taper).- Stated that these standards were followed for evaluation and risk management. No specific results are given for general standards like ISO 14971 (risk management).

    2. Sample size used for the test set and the data provenance

    • Test Set (for non-clinical performance): The document does not specify a "test set" in the context of clinical or diagnostic performance data (e.g., number of patient cases, images). The testing refers to laboratory-based evaluations of the device's material properties, packaging, and stability.
      • Data Provenance: N/A. This is a submission for a new device, and the data provenance refers to materials/engineering tests, not patient data. No country of origin for patient data (as there isn't any provided in text) or retrospective/prospective study type is mentioned as no clinical studies were performed.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • N/A. This information is relevant for studies involving human interpretation or diagnostic accuracy. For this type of device and 510(k) submission, ground truth relates to the results of objective laboratory tests (e.g., is it cytotoxic?), not expert consensus on clinical findings.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • N/A. As no expert consensus or human interpretation of clinical cases was performed, no adjudication method is relevant or described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • N/A. This device is a dental material, not an AI or diagnostic imaging device. Therefore, no MRMC study was conducted or is applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • N/A. This is not an algorithm-based device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for this device is based on objective laboratory test results (e.g., analytical chemistry, toxicology, microbiology, stability) that confirm the material properties, safety, and functionality of the device compared to established scientific principles and the predicate device's known characteristics. It's not clinical outcomes or expert labels.

    8. The sample size for the training set

    • N/A. This term is relevant for machine learning models. No training set is applicable for this type of medical device submission.

    9. How the ground truth for the training set was established

    • N/A. As there is no training set, this question is not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K190220

    Validate with FDA (Live)

    Date Cleared
    2019-06-10

    (125 days)

    Product Code
    Regulation Number
    N/A
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Vista FS and Vista FS Liquid are intended for sulcus retraction prior to impression making and to control bleeding and gingival oozing in restorative and operative dentistry used with gingival retraction cord. Vista FS Liguid facilitate the insertion of the cord into the sulcus.

    Device Description

    Vista FS and Vista FS Liquid are medical devices used to facilitate sulcus retraction prior to taking a dental impressions of a tooth. This entails placement of the sulcus which provides physical displacement of the gingival tissue from the tooth. If using a gingival retraction cord, the subject devices facilitate the insertion of the cord into the sulcus while also facilitating the creating of a physical barrier to prevent gingival bleeding and oozing from affecting restorative and tissue management procedures.

    AI/ML Overview

    The provided text does not contain specific acceptance criteria for performance metrics (such as sensitivity, specificity, or F1 score) or detailed results from a clinical study that would allow a direct comparison to such criteria. Instead, the document focuses on demonstrating substantial equivalence to predicate devices through comparisons of technological characteristics, intended use, and non-clinical performance testing.

    Therefore, many of the requested fields cannot be directly extracted from the provided text. However, based on the information available, here's what can be provided:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative acceptance criteria in terms of clinical performance metrics. The evaluation is based on demonstrating substantial equivalence to predicate devices. The "reported device performance" in this context refers to the results of non-clinical testing confirming manufacturing, cytotoxicity, shelf-life, microbiological properties, and transit resilience.

    Acceptance Criteria (Implicit from Substantial Equivalence Goal)Reported Device Performance
    Manufacturing quality and consistencyTesting verified manufacturing, commensurate with predicate devices.
    Biocompatibility (Cytotoxicity)No differences in cytotoxicity compared to predicate device (ViscoStat Clear); same result at all dilutions.
    Shelf-life stability (interim)All test data for 18-month time point significantly within test acceptance criteria.
    Microbiological safety (bactericidal properties)Exhibit bactericidal properties, mitigating contamination risks and supporting shelf stability.
    Packaging integrity during transitPackaging configurations sufficient, product performed satisfactorily post-transit.
    Clinical performance (equivalent to predicates)Clinical performance is not deemed necessary given substantial equivalence demonstration based on other factors.

    2. Sample size used for the test set and the data provenance

    The document describes non-clinical laboratory testing. It does not refer to a "test set" in the context of clinical data or a patient population. The samples for testing were likely materials from the manufactured devices themselves (e.g., aliquots for analytical testing, cell cultures for cytotoxicity). No information on the country of origin of data or whether it was retrospective or prospective is provided, as these are not relevant to the described non-clinical tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This information is relevant for studies relying on expert-derived ground truth, typically in image analysis or diagnostic scenarios. The provided document concerns non-clinical testing of a dental product.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are used in studies involving human interpretation or subjective assessments, which are not described in this document for the evaluation of this dental product.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a dental product (sulcus retraction material) and not an AI-assisted diagnostic tool.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a dental product and not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the non-clinical tests described would be the established scientific standards and methods for analytical chemistry, cytotoxicity assessment, stability testing, and microbiological evaluation. For example, for cytotoxicity, the "ground truth" is typically determined by observing cell viability and growth inhibition according to standardized protocols (e.g., ISO 10993).

    8. The sample size for the training set

    Not applicable. There is no mention of a "training set" as this document describes non-clinical testing of a physical medical device, not a machine learning model.

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set for a machine learning model, this question is not relevant to the provided document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170101

    Validate with FDA (Live)

    Date Cleared
    2017-06-08

    (148 days)

    Product Code
    Regulation Number
    872.6070
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Source of illumination for curing photo-activated dental restorative materials and adhesives.

    Device Description

    The Valiant Curing Light is a visible light activator for polymerization of dental materials and adhesives, such as dental resins and composites. The Valiant Curing Light is composed of a cordless handpiece, a charging base, disposable barrier sleeves, a light attenuating shield, and an instruction for use. The cordless handpiece is made of medical grade aluminum, electronic elements, glass lenses, and medical grade plastic elements. A plastic rotation switch provides an interface to switch between various modes of the cordless handpiece. Once a mode is selected, an ON/OFF button activates and executes the selected mode. The unit is battery operated and contains a removable and rechargeable Li ion battery. The charging base allows for placement and inductive charging of the handpiece when not in use. The light attenuating shield absorbs emitted light from the cordless handpiece to minimize ocular radiation to the patient and user. The Valiant Curing Light is provided non-sterile. Disposable barrier sleeves (510(k) number K132953, TIDI Shield, Model 21102, by TIDI Products) assist in mitigating patient contamination; additionally, the handpiece, charging base and light attenuating shield can withstand common surface disinfectants. Also included in the Valiant Curing Light is an instructions for use. The instructions for use details the function and four independently operable settings of the device: 1) standard curing mode, 2) ramp curing mode, 3) boost curing mode, and 4) white light mode (visualizing of dental anatomy). The white light mode can be used for intraoral illumination to aid in the visualization of dental anatomy (i.e. equivalent to a class I dental operating light, CFR 872.4630). The curing light modes 1-3 provide broad spectrum blue light (400 - 500mm, with peaks at 410nm and 475nm) to cure a variety of dental restorative materials and adhesives utilizing the most common photoinitiators (CQ and PPD). The white light mode 4 provides an aid for illuminating dental anatomy.

    AI/ML Overview

    The document describes the Valiant Curing Light, a dental device for curing photo-activated dental restorative materials and adhesives. The submission aims to demonstrate substantial equivalence to a predicate device, the Valo Cordless (K110582).

    Here's an analysis of the provided information regarding acceptance criteria and the study that proves the device meets these criteria:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" for performance in a table format with pass/fail thresholds. Instead, it demonstrates performance by comparing the Valiant Curing Light to its predicate device, the Valo Cordless, and asserting "statistically similar" or "equivalent" performance in various non-clinical tests.

    However, based on the non-clinical performance testing sections, we can infer the implied acceptance criteria:

    Acceptance Criteria (Implied from Bench Testing Goals)Reported Device Performance (Valiant Curing Light)
    Composite Hardness: Effectively cure dental composite to depths of 2mm, with statistical similarity to the predicate device.Hardness measurements from cured composite pucks using the Valiant Curing Light were statistically similar to those obtained with the Valo Cordless. Testing confirmed the Valiant adequately cures dental composite to depths of 2mm. (DHF 10018 TR002)
    Wavelength Spectrum: Matches design criteria and is nearly identical to the predicate device.The spectrum emitted from the Valiant matches design criteria. Comparison testing showed both curing lights emit nearly identical wavelength spectrums for curing dental materials. (DHF 10018 TR003)
    Optical Power: Matches design criteria and is identical to or lower than the predicate device.The Valiant's optical power matches design criteria. Comparison testing showed the Valiant has identical or lower optical power results compared to the Valo. The light attenuating shield blocks >99% of emitted optical power. (DHF 10018 TR004)
    Heat Generation: Does not reach unsafe temperatures within typical use.Testing concluded that the Valiant Curing Light did not reach unsafe temperatures within typical use. No hazard or warning statement is required. (DHF 10018 TR001)
    Software Functionality: Software functions properly and complies with FDA guidance.Software was developed in full compliance with the 2005 FDA Software Guidance. Testing verified that the Valiant's software functions properly. (DHF 10018 TR005)
    Electrical Safety & EMC: Complies with relevant standards.The Valiant Curing Light complies with IEC 60601-1 for electrical safety and electromagnetic compatibility.
    Biological Evaluation: Complies with relevant standards.The device complies with ISO 10993-1 (Biological Evaluation of Medical Devices).
    Radiation Safety: Complies with relevant standards (ADA 48, IEC 62471).The device complies with ADA 48 / ISO 10650-2 (radiation safety) and IEC 62471 (Photobiological Safety).

    Study Details

    The document relies solely on non-clinical (benchtop) performance testing to demonstrate substantial equivalence. No clinical studies were deemed necessary.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not explicitly stated for each test (e.g., number of composite pucks, number of light measurements). The descriptions refer to "testing concluded," "measurements obtained," or "comparison testing" without specifying the number of samples or repetitions beyond general terms.
    • Data Provenance: The studies were non-clinical, likely conducted in a laboratory setting by Inter-Med / Vista Dental Products. There is no mention of country of origin of the data as it's not patient data, nor is it retrospective or prospective in the clinical sense.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This question is not applicable as the studies were non-clinical benchtop tests comparing physical properties and performance against a predicate device and engineering design specifications. The ground truth would be established by scientific measurement and engineering validation, not expert consensus in a clinical context.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • This question is not applicable as there were no human readers or expert adjudications involved in establishing ground truth for these non-clinical tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This question is not applicable. This is a dental curing light, not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This question is not applicable. This device is a physical tool (a curing light), not an algorithm. Therefore, "standalone" algorithm performance is not relevant.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for the non-clinical tests was established through objective physical measurements and engineering validation, comparing the device's performance against its own design specifications and against the measured performance of a legally marketed predicate device (Valo Cordless). For example, microhardness measurements would be taken using standardized laboratory techniques.

    8. The sample size for the training set:

    • This question is not applicable. The Valiant Curing Light is a physical device, not an AI algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    • This question is not applicable as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1